Three-Year Outcomes With the Absorb Bioresorbable Scaffold: Individual-Patient-Data Meta-Analysis From the ABSORB Randomized Trials
Three-Year Outcomes With the Absorb Bioresorbable Scaffold: Individual-Patient-Data Meta-Analysis From the ABSORB Randomized Trials
Ali et al., 2018 | Circulation | Meta Analysis
Citation
Ali Ziad A, Gao Runlin, ... Stone Gregg W. Three-Year Outcomes With the Absorb Bioresorbable Scaffold: Individual-Patient-Data Meta-Analysis From the ABSORB Randomized Trials. Circulation. 2018-Jan-30;137(5):464-479. doi:10.1161/CIRCULATIONAHA.117.031843
Abstract
BACKGROUND: The Absorb bioresorbable vascular scaffold (BVS) completely resorbs within 3 years after coronary artery implantation. The safety and effectiveness of BVS through this critical 3-year period have not been characterized. METHODS: We performed an individual-patient-data pooled meta-analysis of the 4 randomized ABSORB trials in which 3389 patients with coronary artery disease were randomly assigned to everolimus-eluting Absorb BVS (n=2164) or cobalt-chromium everolimus-eluting stents (n=1225). The primary efficacy outcome measure was target lesion failure (cardiac mortality, target vessel myocardial infarction, or ischemia-driven target lesion revascularization), and the primary safety outcome measure was device thrombosis. RESULTS: BVS compared with cobalt-chromium everolimus-eluting stents resulted in higher 3-year rates of target lesion failure (11.7% versus 8.1%; risk ratio [RR], 1.38; 95% confidence interval [CI], 1.10-1.73; P=0.006), driven by greater target vessel myocardial infarction (7.8% versus 4.2%; RR, 1.72; 95% CI, 1.26-2.35; P=0.0006) and ischemia-driven target lesion revascularization (6.6% versus 4.4%; RR, 1.44; 95% CI, 1.05-1.98; P=0.02), with comparable cardiac mortality (1.1% versus 1.1%; RR, 0.93; 95% CI, 0.47-1.88; P=0.85). Device thrombosis rates through 3 years were also higher with BVS (2.4% versus 0.6%; RR, 3.71; 95% CI, 1.70-8.11; P=0.001). Between 1 and 3 years, target lesion failure rates (6.1% versus 3.9%; P=0.02) and device thrombosis rates (1.1% versus 0.0%; P<0.0001) were higher with BVS than cobalt-chromium everolimus-eluting stents. CONCLUSIONS: In the present individual-patient-data pooled meta-analysis of the ABSORB trials, BVS was associated with increased rates of target lesion failure and device thrombosis between 1 and 3 years and cumulatively through 3 years of follow-up compared with everolimus-eluting stents. CLINICAL TRIAL REGISTRATION: URL: https://clinicaltrials.gov. Unique identifiers: NCT01751906, NCT01844284, NCT01923740, and NCT01425281.
Key Findings
BVS compared with cobalt-chromium everolimus-eluting stents resulted in higher 3-year rates of target lesion failure (11.7% versus 8.1%; risk ratio [RR], 1.38; 95% confidence interval [CI], 1.10-1.73; P=0.006), driven by greater target vessel myocardial infarction (7.8% versus 4.2%; RR, 1.72; 95% CI, 1.26-2.35; P=0.0006) and ischemia-driven target lesion revascularization (6.6% versus 4.4%; RR, 1.44; 95% CI, 1.05-1.98; P=0.02), with comparable cardiac mortality (1.1% versus 1.1%; RR, 0.93; 95% C
Outcomes Measured
- Requires manual extraction
Population
| Field | Value |
|---|---|
| Population | coronary artery disease were |
| Sample Size | 2164 |
| Age Range | See abstract |
| Condition | See abstract |
MeSH Terms
- Absorbable Implants
- Coronary Artery Disease
- Coronary Thrombosis
- Drug-Eluting Stents
- Humans
- Myocardial Infarction
- Percutaneous Coronary Intervention
- Prosthesis Design
- Randomized Controlled Trials as Topic
- Recurrence
- Risk Factors
- Time Factors
- Treatment Outcome
Evidence Classification
- Level: Meta Analysis
- Publication Types: Journal Article, Meta-Analysis, Research Support, Non-U.S. Gov't
- Vertical: chromium
Provenance
- PMID: 29089314
- DOI: 10.1161/CIRCULATIONAHA.117.031843
- PMCID: Not in PMC
- Verified: 2026-04-09 via PubMed E-utilities API
Source extracted via PubMed E-utilities API on 2026-04-09