A Systematic Review of Systemic Cobaltism After Wear or Corrosion of Chrome-Cobalt Hip Implants
A Systematic Review of Systemic Cobaltism After Wear or Corrosion of Chrome-Cobalt Hip Implants
Gessner et al., 2019 | J Patient Saf | Systematic Review
Citation
Gessner Bradford D, Steck Thomas, ... Tower Stephen S. A Systematic Review of Systemic Cobaltism After Wear or Corrosion of Chrome-Cobalt Hip Implants. J Patient Saf. 2019-Jun;15(2):97-104. doi:10.1097/PTS.0000000000000220
Abstract
OBJECTIVES: We sought to synthesize data on systemic arthroprosthetic cobaltism, a recently described syndrome that results from wear or corrosion of chrome-cobalt hip components. METHODS: We conducted a systematic literature review to identify all reported cases of systemic arthroprosthetic cobaltism. To assess the epidemiologic link between blood cobalt levels (B[Co]), we developed a symptom scoring tool that evaluated 9 different symptom categories and a category of medical utilization. RESULTS: We identified 25 patients reported between 2001 and 2014 with a substantial increase in case reports over the past 3 years. Symptoms were diverse and involved the hip (84%), cardiovascular system (60%), audiovestibular system (52%), peripheral motor-sensory system (48%), thyroid (48%), psychological functioning (32%), visual system (32%), and the hematological, oncological, or immune system (20%). The mean latency from implantation to presentation or revision was 41 months (range, 9-99 months). The mean B[Co] was 324 μg/L and 4 patients had levels less than 20 μg/L. The B[Co] but not blood chromium level was highly associated with a quantitative measure of overall symptom severity (r, 0.81; P < 0.001). Mean B[Co] and symptom scores were substantially higher in patients with revisions of failed ceramic-on-ceramic prostheses than those with primary metal-on-metal prostheses. CONCLUSIONS: Systemic arthroprosthetic cobaltism is an increasingly recognized complication of wear or corrosion of chrome-cobalt hip implants, may involve a large number of organ systems, and may occur with relatively low B[Co]. There is an urgent need to better define the overall scope of the problem and to develop screening and management strategies.
Key Findings
We identified 25 patients reported between 2001 and 2014 with a substantial increase in case reports over the past 3 years. Symptoms were diverse and involved the hip (84%), cardiovascular system (60%), audiovestibular system (52%), peripheral motor-sensory system (48%), thyroid (48%), psychological functioning (32%), visual system (32%), and the hematological, oncological, or immune system (20%). The mean latency from implantation to presentation or revision was 41 months (range, 9-99 months).
Outcomes Measured
- Requires manual extraction
Population
| Field | Value |
|---|---|
| Population | revisions of failed ceramic |
| Sample Size | 25 |
| Age Range | See abstract |
| Condition | See abstract |
MeSH Terms
- Adult
- Aged
- Arthroplasty, Replacement, Hip
- Cobalt
- Corrosion
- Female
- Hip Prosthesis
- Humans
- Male
- Middle Aged
Evidence Classification
- Level: Systematic Review
- Publication Types: Journal Article, Research Support, Non-U.S. Gov't, Systematic Review
- Vertical: chromium
Provenance
- PMID: 26076080
- DOI: 10.1097/PTS.0000000000000220
- PMCID: PMC6553976
- Verified: 2026-04-09 via PubMed E-utilities API
Source extracted via PubMed E-utilities API on 2026-04-09